DURATEC Analysentechnik GmbH
Flowmeter PDF Print E-mail
Monday, 16 April 2007 19:17

vflowsmallThe liquid flowmeter vFlow measures the real flow rate of a liquid stream. It determines the required time for filling a measuring tube with a constant volume. The LCD display shows the measured value and the filling status. As soon as a new flow rate is calculated, the value  is sent via RS-232-interface to a computer or printer as data logger.

Datasheet (View / Download )

The vFlow is especially designed for performance tests of HPLC or metering pumps. The accuracy and fluctuation of the flow can be monitored. The flowmeter can be calibrated according to NIST or DKD traceable standards. Materials for wetted parts are PEEK, PTFE or glass which make the device  resistant to aggressive media. The delivery includes power adapter, mounting bracket and certificate of calibration.